GENERAL FEATURES
- High-contrast 4.3-inch color LCD touchscreen display.
- An intelligent operating system includes features such as:
- User management;
- Calibration management;
- Electrode management;
- Method management;
- Data management;
- Log management.
- Multi-reading modes: auto-read, timed-read, and continuous-read.
- Automatic and manual temperature compensation for accurate results.
- Auto-hold function automatically senses and locks the measurement endpoint.
- Data storage: up to 1000 sets per parameter (GLP-compliant).
- Built-in data analysis function for reviewing, comparing, and recalculating results.
- USB communication supported.
- Auto power-off feature helps extend battery life.
- The reset function restores all settings to factory defaults.
- IP65-rated waterproof design suitable for field and outdoor measurements.
Category |
Specification |
Model |
M510T-A |
Dimensions |
90 × 255 × 40 mm |
Weight |
500 g (1.1 lb) |
Power |
Rechargeable Lithium Battery, AC Adapter (100–240V AC input, DC 5V output) |
Backlight |
Yes |
Measured Parameters |
pH, EC, ISE, Temperature, mV/ORP, pX, Resistivity, TDS, Salinity |
Connector |
Type |
pH Electrode |
BNC (Q9) |
DO with Temp. Probe |
4-pin aviation connector |
Conductivity Probe |
5-pin aviation connector |
Package Includes
- E-301-QC 3-in-1 pH composite electrode
- DJS-1VTC EC electrode
- NIST pH standard solutions (pH 4.01, 7.00, 10.01), 50 mL/vial
- 1413 μS/cm EC standard solution, 50 mL, 1 vial
- Electrode holder
- Silicone rubber protective case
- Wristbands
- Touch pen
- Carrying case
pH Measurement
- 1 to 8-point calibration with standard buffer recognition.
- Selectable pH buffer groups: USA, NIST, DIN, GB, MERK, JIS.
- Supports user-defined pH buffers or buffer groups.
- Automatic electrode diagnostics with slope and offset display.
Property |
Details |
Range |
-2.000 to 20.000 pH |
Resolution |
0.1, 0.01, 0.001 pH |
Accuracy |
±0.002 pH |
Calibration Points |
Up to 8 |
Standard Customization |
Yes |
Calibration Reminder |
Yes |
Standard Recognition |
USA, NIST, MERK, JIS, GB, DIN |
Slope Limit |
Yes |
Ion Measurement
- 1 to 6-point calibration.
- Selectable units: μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.
- Supports multiple measurement modes:
- Direct Reading
- Standard Addition
- Sample Addition
- Gran Plot
- Built-in methods for over 10 common ions, including: F⁻, Cl⁻, Br⁻, I⁻, NO₃⁻, BF₄⁻, NH₄⁺, K⁺, Na⁺, Ca²⁺, Cu²⁺, Pb²⁺, Ag⁺, etc.
- User-defined methods are also supported.
Property |
Details |
Range |
1.000e-9 to 9.999e+9 |
Units |
mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb |
Resolution |
Up to 4 significant digits |
Accuracy |
±0.3% |
Calibration Points |
Up to 8 |
Conductivity Measurement
- 1 to 5-point calibration with automatic recognition of standard solutions.
- Configurable parameters:
- Cell constant
- Reference temperature (5/10/15/18/20/25°C)
- Temperature compensation coefficient
- TDS conversion factor
- Temperature compensation types: none, linear, pure water.
- Range: 0.000 μS/cm to 3000 mS/cm
- Resolution: 0.001 μS/cm minimum